Contra-indications of 3-S

ecat

Herewith the replies to my query >Anyone able to list a few specific

contraindications for 3-s system or at

>least point me at a website where I can read same please? Thankyou to all

who responded. This kind of feedback does assist a bloke to make his own

decisions based on the experiences of his colleagues.

>Richard Ziegeler (P&O)

I have been using the 3/S system using custom Durr-Fillaur locking system

for over 10 years. It helps the limb heal, protects the limb from improper

donning, and when the patient does not following instruction. My old

patients would die if asked them to use socks against their skin. We even

use it on most AKA. There are some tricks to having it work, but it is

pretty easy. Socks cause revisions and abrasions that don’t need to

happen.

[email protected]

John G. Russell Jr.

3161 Putnam Blvd.

Pleasant Hill, CA. 94523

Phone 510-943-1119

Fax 510-943-24-93

Dear Mr. Ziegeler,

Some of the contraindications for 3-S that I can recall are:

1) Upper Extremity Involvement such as severe arthritis or upper extremity

prosthetic use

2) Excessive Perspiration that does not dissapate over time of wear

3) Unacceptance of distal end “milking” action of the residuum

4) Skin that breaks down easily with exposure to silicone

5) Deeply invaginated scars or muscle bunching that can break suction seal

6) Highly conical, bony, and short residual limb shape

I am sure there are more but these might give you a start.

Sincerely,

Gerald Stark, BSME,CP

Fillauer, Inc.

I have had a lot of success with the Fillauer 3-S and Ossur Icex systems.

Some things to watch out for are invaginated scars, adherent tissue or

other skin sensitivity. Conical or extremely prominent tibia can pose a

problem if the

correct system is not used. I would recommend the Iceross liner w/distal

cup for many of these situations. Hydrostatic or total surface bearing

sockets work well for most. Use your judgment or call for technical support

at Ossur.

Speak with either Hans Schaeffer or Jan (Yan) Christenson. Hope this helps.

Ryan Gatlin

Hi Richard,

By the 3-system do you mean the silicone liners in themselves or the other

parts of the system such as the locking method (ie

shuttle-lock/lanyard/loop and pin)?

Also do you mean this particular system by this manufacturer (about which I

know next to nothing) or does the question apply equally to ALPS or Ossur

systems (both of which I have used a fair bit)?I’ll be happy to help if I

can.

Cheers,

Grant Crosthwaite

Richard Some off-the-top-of-my-head contraindications for 3 S system:

VERY low tolerance to distal contact

Mobile, redundant distal tissue (difficulty donning)

Upper extremity limitations (poor manual dexterity, arthritis or

amputation)

No room for the shuttle lock (long residual limb)

Allergy to silicone or other sleeve materials

Open residual limb wound

Poor hygeine habits

High cost for purchase and/or replacement

Some long time sock wearers just don’t like the feel of the gel sleeves

I hope this smattering of information/experience is helpful. Good luck!

Dave ———————————-

DMProcter, CPO

OmniCare Labs

Some say that people with compromised skin, say envaginated scars or burn

scars, may not be candidates for 3-S-type systems. I think proper socket

fit and use of the liner best for the person, eg OWW Alpha, ICEROSS

Comfort Plus, the new TEC with pin, etc, is key.

Some complain of distal pistoning. That results from liners chosen too big

for the residuum, lack of an air seal between socket and liner (a

semi-hypobaric fit on outside of the liner as well as inside is a real

plus), or improper socket fit. As for the air seal, I believe the latches

by Ortho Europe (see www.ortholite.co.uk), which I think are the best, are

now supplied with o-ring seals on the latch pins for semi-hypobaric fit.

I’m working to confirm that….

I personally favor the thicker liners, like the 9mm Alpha and the thick

Comfort Plus. Thin liners, like the old ICEROSS and Aegis have too little

padding and aren’t as forgiving of volume changes. I use both a full

length and shortened Alpha 9mm Uniform liners. Talk about cushioning!

Chris Johnson, Engineering Director,

College Park Industries, Inc.

[email protected]

www.college-park.com

 

RECENT NEWS

Get unlimited access!

Join EDGE ADVANTAGE and unlock The O&P EDGE's vast library of archived content.

O&P JOBS

The O&P EDGE Magazine
Are you sure want to unlock this post?
Unlock left : 0
Are you sure want to cancel subscription?

Welcome Back!

Login to your account below

Retrieve your password

Please enter your username or email address to reset your password.

The O&P EDGE Magazine
Are you sure want to unlock this post?
Unlock left : 0
Are you sure want to cancel subscription?